Troubleshoot - Harvey - Hydroclave

Message What To Do
SELF TEST FAILED CALL SERVICE Call for authorized service.
  Press OFF, then ON. If "CONTROLS ERROR" message is not cleared, go to step #2.
CONTROLS ERROR CALL SERVICE Unplug the unit, then press and hold OFF button while plugging back in. Release OFF button after "PERFORMING SELF-TEST" is displayed. If step #2 does not clear the "CONTROLS ERROR" message, call for authorized service.
HEAT UP TOO LONG CYCLE CANCELED REPROCESS LOAD Reprocess load and run cycle again. Call service if message occurs two cycles in a row.
OVERHEAT CYCLE CANCELED REPROCESS LOAD Check water level in reservoir. Remove and clean reservoir and chamber screens. Reprocess load and run cycle again. Call service if message occurs two cycles in a row.
VENT TOO LONG CHECK FOR DEBRIS IN CHAMBER Wait for chamber to cool, then open door. Check for debris in chamber or clean chamber filter. Reprocess load and run cycle again. Call service if message occurs two cycles in a row.
DOOR OPEN CYCLE CANCELED REPROCESS LOAD Be sure door is closed and latched. (Press down.) Reprocess load and run cycle again.
PRESSURE TOO LOW CYCLE CANCELED REPROCESS LOAD Clean gasket and check for rips and holes. Reprocess load and run cycle again. Call service if message occurs two cycles in a row.
TEMPERATURE TOO LOW CYCLE CANCELED REPROCESS LOAD Reprocess load and run cycle again. Call service if message occurs two cycles in a row.
UNIT ERROR PRESS OFF TO CONTINUE Press OFF, then press ON again.
PRESSURE TOO HIGH CYCLE CANCELED REPROCESS LOAD Reprocess load and run cycle again. Call service if message occurs two cycles in a row.

 

Problem Possible Cause What To Do
Wrapped loads are wet at completion of sterilize cycle. Load improperly loaded in chamber. Be sure wrap is not touching the walls of the chamber. Do not place wrapped goods on bottom of tray rack (10" chamber only).
Load does not dry in drying cycle. Inadequate drying time. Sterilizer door not opened soon enough after completion of sterilize cycle. Load excessively wet: • Load size too large • Load touching chamber walls • Wrapped load in bottom of tray rack (10" chamber only). Door open too far or not far enough. Increase drying time. Open the door immediately after completion of a cycle. Refer to "Load Chart" and "Loading" procedures. Door should be open approximately 1/4" if using open door drying method. (Push latch down while door is open.)

 